User Tools

Site Tools


teamc2024

This is an old revision of the document!


Team C - PMT (drafting)

Group members:

  • Mohan Karthik, Tata Inst., IN
  • Nguyen Gia Minh, Fulbright Univ., HCM
  • Ta Van Minh, HUS-VNU, HN
  • Hoang Pham Thanh Dat, HUS-VNU, HCM

Mentor: Dr. Son Cao

Metal-package PMT module: H6780-1 / backup H10721-110 (newly purchased, using for O-BLM and for other research)

  • Voltage apply (tested) ~ 14V (range 11.5V to 15.5V)
  • Gain driving (<1V)/ tested 0.9V
  • Peak wavelength sensitivity is around 400nm

Three main objectives

  • PMT noise rate, raising, falling; PMT vs. siPM (quantum efficiency, electric gain, linearity, dynamic range)
  • A concept of fast spectrometer: Get spectra of a white light source with optical filter and PMT
  • Cosmic ray muons measurement (rate, light yield) with plastic scintillator cubes

Important notes

  • PMT voltage supply <15V and for gain control <1V
  • MPPC voltage supply <60V
    • Some RF noise is induced by this voltage supply. If you are using PMT only, do not turn o this supply.
  • Light tightening is supper important for PMT (than for MPPC).
  • Be gentle when dealing with optical fibers

Allocated hardwares

Reference